MATRIX 200™ REFERENCE MANUAL
8
8.4.3 Calibration
VisiSet™ provides a Calibration Tool to maximize the reading performance by tuning the
acquisition parameters and the time of the delayed triggers.
By selecting the Calibration Tool from the VisiSet™ Main Menu (F), the following window
appears together with the Parameter Setup window:
Figure 104 - Calibration OK
This tool provides a "real-time" image display while Matrix 200™ is reading. It also gives
immediate results on the performance of the installed Matrix 200™ reader.
The Parameter Setup window works in Interactive Mode in order to cause each parameter
setting to be immediately effective.
NOTE
If you want to save the temporary configuration to permanent memory, you
must first close the Calibration Tool window. Then, you must disable the
Interactive Mode and select the Permanent Memory option from the Send
Configuration item in the Device menu.
94